Accu-Chek Performa
Error codes E-6
Blood or control solution was applied to the test strip before the flashing drop symbol appeared on the display. Discard the test strip and repeat the blood glucose or control test.
Blood or control solution was applied to the test strip before the flashing drop symbol appeared on the display. Discard the test strip and repeat the blood glucose or control test.
Get the most out of your Accu-Chek products with support and troubleshooting tips.
Not enough blood or control solution was drawn into the test strip for measurement or was applied after the test had started. Discard the test strip and repeat the blood glucose or control test.
Get the most out of your Accu-Chek products with support and troubleshooting tips.
Your blood glucose may be extremely high or a meter or a test strip error has occurred.
Get the most out of your Accu-Chek products with support and troubleshooting tips.
The test strip may be damaged or not properly inserted. Remove and reinsert the test strip, or replace it if damaged.
Get the most out of your Accu-Chek products with support and troubleshooting tips.
Your meter measures blood glucose within a defined range (0.6–33.3 mmol/L).
The display may indicate that your blood glucose is lower than the measuring range of the system (0.6 mmol/L).
Immediately repeat the test. If your test result matches how you feel, contact your healthcare professional immediately.
If this does not match how you feel, contact (insert affiliate contact information)
Get the most out of your Accu-Chek products with support and troubleshooting tips.
Your meter measures blood glucose within a defined range (0.6–33.3 mmol/L).
The display may indicate that your blood glucose is above than the measuring range of the system (33.3 mmol/L).
Immediately repeat the test. If your test result matches how you feel, contact your healthcare professional immediately.
If this does not match how you feel, contact (insert affiliate contact information)
Get the most out of your Accu-Chek products with support and troubleshooting tips.
Battery is dead. Insert new battery. Insert one 3-volt lithium battery (coin cell type CR2032)
Display is damaged. Contact (insert affiliate contact information)
Meter is defective. Contact (insert affiliate contact information)
Extreme temperatures. Move the meter to a more temperate area.
Get the most out of your Accu-Chek products with support and troubleshooting tips.
To obtain a blood sample, the AST cap is pressed against the skin on the punctured site. The bottom of the cap is transparent to enable visibility of puncture site when pressing.
Check other FAQ categories by clicking on the below category names.
No. It is intended to be used by one person only and shall not be used on multiple users. If someone else uses your lancing device there is an increased risk of cross infections.
Check other FAQ categories by clicking on the below category names.
No. It is not designed for use on pets.
Check other FAQ categories by clicking on the below category names.